◤OVERVIEW
The HOBO Water Level Titanium is recommended for saltwater deployment for recording water levels and temperatures in wetlands and tidal areas. This data logger features high accuracy at a great price and HOBO ease-of-use, with no cumbersome vent tubes or desiccants to maintain.
Helpful Links:
。Barometric Pressure Compensation Assistant Demo
- Lightning protection – no long signal wires, and electronics are shielded in a titanium housing for use in saltwater (see the Water Level logger sensor location drawing)
- HOBOware Pro software provides easy conversion to accurate water level reading, fully compensated for barometric pressure (see demo) temperature, and water density.
- Multiple-rate sampling (see demo) allows faster sampling at critical times such as when pumping starts or stops.
- Also available in narrow depth range (U20-001-04-Ti)
- Ideal for use in wells, streams, lakes, wetlands and tidal areas
- No-vent-tube design for easy reliable deployment
- Available in stainless and titanium versions
- Durable ceramic pressure sensor
- 3-point NIST-traceable calibration certificate included
In what environment does this data logger operate?
This data logger operates in outdoor and underwater environments.
What measurements does this data logger support?
The U20-001-01-Ti data logger supports the following measurements: Absolute Pressure, Barometric Pressure, Temperature and Water Level
Pressure and Water Level Measurements U20-001-01 and U20-001-01-Ti
- Operation Range:0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
- Factory Calibrated Range:69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
- Burst Pressure:310 kPa (45 psia) or 18 m (60 ft) depth
- Water Level Accuracy*:
。Typical error:±0.05% FS, 0.5 cm (0.015 ft) water
。Maximum error:±0.1% FS, 1.0 cm (0.03 ft) water - Raw Pressure Accuracy**:±0.3% FS, 0.62 kPa (0.09 psi) maximum error
- Resolution:<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
- Pressure Response Time (90%)***:<1 second; measurement accuracy also depends on temperature response time
- Operation Range:-20° to 50°C (-4° to 122°F)
- Accuracy:±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A in manual
- Resolution:0.10°C at 25°C (0.18°F at 77°F), see Plot A in manual
- Response Time (90%):5 minutes in water (typical)
- Stability (Drift):0.1°C (0.18°F) per year
- Real-time Clock:± 1 minute per month 0° to 50°C (32° to 122°F)
- Battery:2/3 AA, 3.6 Volt lithium, factory-replaceable
- Battery Life (Typical Use):5 years with 1 minute or greater logging interval
- Memory (Non-volatile):64K bytes memory (approx. 21,700 pressure and temperature samples)
- Weight:
。Stainless steel models:approximately 210 g (7.4 oz)
。Titanium models:approximately 140 g (4.8 oz) - Dimensions:2.46 cm (0.97 inches) diameter, 15 cm (5.9 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
- Wetted Materials:Titanium, Viton® o-rings, acetyl cap, ceramic sensor
- Logging Interval:Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware software manual.
- Launch Modes:Immediate start and delayed start
- Offload Modes:Offload while logging; stop and offload
- Battery Indication:Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile.
- Environmental Rating:IP68
- :The CE Marking identifies this product as complying with all relevant directives in the European Union (EU).
* Water Level Accuracy:With accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data, and a stable temperature environment.
** Raw Pressure Accuracy:Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors.
*** Changes in Temperature:Allow 10 minutes in water to achieve full temperature compensation of the pressure sensor. Maximum error due to rapid thermal changes is approximately 0.5%.