◤OVERVIEW
The Type T 6 ft Beaded Thermocouple probe includes 1.8m (6′) of insulated 30-AWG wire wound on an integrated spool caddy /subminiature connector. Color-coded for type T (blue) thermocouples according to the ANSI standard, this probe is teflon-insulated and responsive over a range of -200° to 180°C (-330° to 356°F) with an initial accuracy of ±1.5°C (±2.7°F). Overall accuracy varies with attached logger.
- 30-AWG wire
- Teflon-insulated probe
In what environment does this sensor operate?
This sensor operates in an indoor environment.
What measurements does this sensor support?
The TC6-T sensor supports the following measurement: Temperature
- Range & Accuracy:
。-200 to 65°C (-330° to -85°F) – ±1.5°% accuracy
。-65 to 130°C (-85° to 270°F) – ±1.0°C (± 1.8°F) accuracy
。130 to 180°C (270° to 356°F) – ±0.75% accuracy - Insulation: Type T Teflon®
*Note: Teflon insulation is rated up to 200°C (400°F) continuous, and +250°C (+482°F) for 24 hours.
**Note: Percentage is in degrees above or below ice point (0°C or 32°F)